Found 2 results

510(k) Data Aggregation

    K Number
    K102173

    Validate with FDA (Live)

    Date Cleared
    2010-09-13

    (42 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The model 1712 Moldex-Metric Type N95 Healthcare Particulate Respirator and Surgical Mask meets CDC Guidelines for TB Exposure Control within healthcare facilities. This device is also intended to be worn by healthcare personnel during surgical procedures to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material.

    Device Description

    The Moldex-Metric type N95 Healthcare Particulate Respirators and Surgical Masks are constructed from an extruded plastic mesh used in the outer cover and a nonwoven spunbond used in the inner and outer cover. The polypropylene melt blown filter media is layered between the inner and outer cover. The head strap is made of a non-latex rubber stapled to the mask. The inside nosepiece is a closed cell foam.

    AI/ML Overview

    The provided document is a 510(k) summary for the Moldex-Metric Model 1712 N95 Healthcare Particulate Respirator and Surgical Mask. It describes the device, its intended use, and performance tests conducted to demonstrate substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicTest MethodAcceptance CriteriaDevice Performance (Model 1712)
    Fluid ResistanceASTM 1862 - 00a32 of 32 pass32 of 32 pass
    Flammability16 CFR 1610Flame spread must be within upper and lower limits / No flame spread on 10 of 10 samples, meets Class INo flame spread on 10 of 10 samples, meets Class I
    Filter Efficiency (%)NIOSH, 42 CFR Part 84> 95% EfficientAverage 98.95% efficient of 20 samples
    Inhalation Breathing Resistance (mm H2O)NIOSH, 42 CFR Part 84< 35.0 mm H2O @ 85 lpmAverage 8.9 mm H2O @ 85 lpm of 3 samples
    Exhalation Breathing Resistance (mm H2O)NIOSH, 42 CFR Part 84< 25.0 mm H2O @ 85 lpmAverage 8.9 mm H2O @ 85 lpm of 3 samples
    Biocompatibility (Cytotoxicity)ISO 10993-1Score of 2 or lessScore of 0
    Biocompatibility (Sensitization)ISO 10993-1Grade 1 (no different than control)Grade 1
    Biocompatibility (Primary Skin Irritation)ISO 10993-1NegligibleNegligible
    Bacterial Filtration EfficiencyModified Greene and Vesley Method> 99.9%> 99.9%

    2. Sample Size Used for the Test Set and Data Provenance

    • Fluid Resistance: 32 samples were tested. Data provenance is not explicitly stated (e.g., country of origin), but the test was conducted in accordance with ASTM F 1862, a recognized international standard. The test appears to be prospective for the Model 1712.
    • Flammability: 10 samples were tested. Data provenance is not explicitly stated, but the test was conducted in accordance with 16 CFR 1610. The test appears to be prospective for the Model 1712.
    • Filter Efficiency: 20 samples were tested. Data provenance is not explicitly stated, but the test was conducted by NIOSH in accordance with 42 CFR Part 84, suggesting a U.S. regulatory context. The test appears to be prospective; the device was certified by NIOSH.
    • Breathing Resistance (Inhalation and Exhalation): 3 samples were tested for each. Data provenance is not explicitly stated, but the test was conducted by NIOSH in accordance with 42 CFR Part 84, suggesting a U.S. regulatory context. The test appears to be prospective; the device was certified by NIOSH.
    • Biocompatibility (Cytotoxicity, Sensitization, Primary Skin Irritation): The document states these tests were conducted on predicate devices made from the same material. The specific number of samples for these predicate device tests is not detailed in this summary.
    • Bacterial Filtration Efficiency: The document states this test was conducted using the Modified Greene and Vesley Method. The specific sample size is not detailed for the Model 1712 directly, but it indicates "Greater than 99.9%" results derived from "Test results show a bacterial filtration efficiency greater than 99.9%." It also notes that this test was "conducted using the Modified Green and Vesley Method... for the predicate devices."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This document describes the performance testing of a physical device (respirator mask) against established technical standards and regulatory requirements. It does not involve human expert interpretation of medical images or data where "ground truth" is established by medical experts in the traditional sense (e.g., radiologists interpreting images). The "ground truth" here is objective measurement against specified physical and biological performance standards.

    4. Adjudication Method for the Test Set

    Not applicable. The tests involve objective measurements and compliance with numerical or categorical standards (e.g., "32 of 32 pass," ">95% efficient," "Score of 0"). There is no subjective interpretation requiring an adjudication process for consensus among experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a physical product (respirator mask), not an AI algorithm for diagnostic interpretation that would involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Yes, the device's performance was evaluated in a standalone manner against objective performance criteria. The tests (fluid resistance, flammability, filter efficiency, breathing resistance, biocompatibility, bacterial filtration efficiency) assess the physical and biological characteristics of the mask itself, without human-in-the-loop performance being a variable in these specific tests.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests described are:

    • Objective physical and biological measurements: This includes quantified values (e.g., percentage efficiency, pressure measurements in mm H2O) and qualitative assessments (e.g., pass/fail for fluid resistance, flame spread classification, biocompatibility scores) against established industry and regulatory standards (ASTM, NIOSH, CFR, ISO).
    • Predicate device performance data: For some tests (like biocompatibility and bacterial filtration efficiency), the ground truth for the new device's substantial equivalence is established by demonstrating that it performs similarly to or better than a legally marketed predicate device, which itself met regulatory ground truth requirements.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, and the testing described does not involve machine learning or AI that would require a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062663

    Validate with FDA (Live)

    Date Cleared
    2006-10-11

    (34 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The various models of Moldex Type N95 Healthcare Particulate Respirators and Surgical Masks meet CDC Guidelines for TB Exposure Control within healthcare facilities. These devices are also intended to be worn by healthcare personnel during surgical procedures to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material.

    Device Description

    The Moldex type N95 Healthcare Particulate Respirators and Surgical Masks are constructed from an extruded plastic mesh used in the outer cover and a nonwoven spunbond used in the inner and outer cover. The polypropylene melt blown fitter media is layered between the inner and outer cover. The head straps are made of a non-latex rubber stapled to the mask, The inside nosepiece is a closed cell foam.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Moldex Health Care N95 Particulate Respirators and Surgical Masks, based on the provided text:

    Acceptance Criteria and Device Performance

    Performance CharacteristicsTest MethodAcceptance CriteriaReported Device Performance (Moldex Health Care N95 Particulate Respirators and Surgical Masks)
    Fluid Resistance PerformanceASTM 1862-00a32 of 32 pass32 of 32 pass
    Flammability Class16 CFR 1610Flame spread must be within upper and lower limits / Meets Class 1No flame spread on 10 of 10 samples, meets Class I
    Filter Efficiency (%)NIOSH, 42 CFR Part 84> 95% EfficientAverage 96.81% efficient of 20 samples
    Breathing Resistance (mm H₂O)NIOSH, 42 CFR Part 84< 35.0 mm H₂O @ 85 lpmAverage 10.2 mm H₂O @ 85 lpm of 3 samples
    Biocompatibility (Cytotoxicity)ISO 10993-5 (tested on predicate devices)Score of 2 or lessSame as predicate device (Predicate device score of 0, so Moldex also meets this)
    Biocompatibility (Sensitization)ISO 10993-10 (tested on predicate devices)Grade 1 (no different than control)Same as predicate device (Predicate device Grade 1, so Moldex also meets this)
    Biocompatibility (Primary Skin Irritation)ISO 10993-10 (tested on predicate devices)NegligibleSame as predicate device (Predicate device Irritation Negligible, so Moldex also meets this)
    Bacterial Filtration EfficiencyModified Greene and Vesley Method. J Bacteriol 83:663-667.Greater than 99.9%Test results show a bacterial filtration efficiency greater than 99.9%

    Study Details

    1. Sample sizes used for the test set and data provenance:

      • Fluid Resistance Performance: 32 samples (number of individual masks tested). Data provenance is not specified, but typically these are laboratory tests performed on samples from production batches.
      • Flammability Class: 10 samples. Data provenance is not specified.
      • Filter Efficiency (%): 20 samples. Data provenance is not specified.
      • Breathing Resistance (mm H₂O): 3 samples. Data provenance is not specified.
      • Biocompatibility (Cytotoxicity, Sensitization, Primary Skin Irritation): Tested on unspecified number of predicate devices. Data provenance is not specified.
      • Bacterial Filtration Efficiency: Not explicitly stated, but the test results show "a bacterial filtration efficiency greater than 99.9%". This would involve testing multiple samples. Data provenance is not specified.

      All tests are laboratory-based performance evaluations, not clinical studies involving human subjects in particular countries. They are inherently prospective tests performed on samples of the device to demonstrate compliance with standards.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This report describes the results of physical and performance testing of a medical device (respirators/surgical masks) against established industry standards and regulatory requirements. It does not involve "experts" establishing a "ground truth" in the way one might for diagnostic imaging interpretation. The "ground truth" is defined by the objective measurement criteria of the specified test methods (e.g., ASTM F1862 for fluid resistance, NIOSH 42 CFR Part 84 for filter efficiency). The laboratories performing these tests are accredited for these specific testing methodologies, and their expertise lies in conducting these technical measurements accurately according to protocol.

    3. Adjudication method for the test set:
      Not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where human interpretation or subjective assessment is a component of ground truth determination (e.g., grading medical images). This document describes objective performance testing.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This is not a study involving human readers or AI assistance; it's a device performance and safety evaluation.

    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
      Not applicable. This document pertains to physical device performance, not algorithm performance.

    6. The type of ground truth used:
      The "ground truth" is defined by the objective, quantitative measurements and pass/fail criteria established by standardized test methods (e.g., ASTM, NIOSH, ISO). For instance, for filter efficiency, the ground truth is simply whether the filter achieved >95% efficiency when tested according to NIOSH 42 CFR Part 84.

    7. The sample size for the training set:
      Not applicable. This document describes the testing of finished products, not the development or training of an algorithm.

    8. How the ground truth for the training set was established:
      Not applicable. There is no algorithm training set discussed in this document.